A Technique of Efficiency Calibration of HPGe Detector

Authors

  • Saeid Y. Elorfi
  • Marai M. Imsallim
  • Yasin K. Abdalla

DOI:

https://doi.org/10.37376/ljst.v7i2.3165

Keywords:

HPGe Detector; Calibration; Efficiency

Abstract

In this work, a technique for the determination of efficiency calibration of hyper-pure germa-
nium detector was obtained using four different concentrations of KCl in Bq/l for different

geometrical configurations. The relative intensities of Gamma-rays from 226Ra with its Gamma-
Emitting daughters were used in addition to the normalizing factor relative to the intensities of

gamma-rays photo peaks. The results of this method were compared to standard samples of
IAEA-373 and they were found to be in good agreement.